By adding gas exchange for the direct measurement
of VO2, the CPX Ultima provides objective, reproducible
data on functional capacity, along with a method for determining
the primary cause of exercise intolerance or dyspnea. Standard
stress and resting pulmonary function tests cannot measure
the complete cardiorespiratory system. The CPX Ultima makes
these measurements and more, with unparalleled accuracy.
With a completely digital data acquisition and precise
breath by breath analysis, the CPX Ultima sets the benchmark
for gas exchange systems in the world today.
Combined with the patented MedGraphics preVent™
Pneumotach, your testing will be free of cumbersome hoses
and headgear. After testing, you can change the preVent™
Pneumotach in a matter of seconds, giving you an instant solution
for infection control.
BreezeSuite Software
BreezeSuite software combines easy to use, powerful data collection
with data review on a single screen. The software comes with
integrated features to enable custom configurations and future
expandability, along with an open architecture database to
provide instant data access. Helpful coaching features walk
the user through simple steps for systems calibration and
patient testing. BreezeSuite allows you to design custom reports
and configurations, using Report and Script Designer features.
Physician Review Station features full trending of all variables
for enhanced patient management. BreezeSuite is compatible
with Windows 2000™ and Windows XP Professional™. |



All specifications
are subject to change without notice. Products may vary
from those illustrated. Please contact your Medical Graphics
representative for the latest information, pricing and product
availability.
©
2004-2006 Medical Graphics Corporation. CPX Ultima and
preVent are registered trademarks of Medical Graphics
Corporation.
|
Specifications:
Innovative Technology
- Patented breath-by-breath analysis and rapid response
O2 and CO2 analyzers.
- Patented, lightweight (26 gm/<1 oz.) preVent pneumotach,
exceeds ATS/ERS performance standards and eliminates the
need for bulky headgear and cumbersome connections.
- Automated detection of anaerobic threshold, respiratory
compensation and VO2max.
- Five additional inputs for monitoring pulse-oximetry,
heart rate, blood pressure, and outputs from treadmills
and cycles.
Flow Device
- Bi-directional differential pressure preVent
pneumotach
- U.S. Patent Numbers 5038773 & 5119825
- Range: ± 18 L/sec
- Accuracy: ± 3% or 50 ml (meets or exceeds ATS/ERS clinical
performance standards)
- Resolution: 8.64 ml/sec
- Deadspace: 39 ml
Gas Sample
- Patent Number 5042500
- Patented gas drying sample circuit
- Side stream sampling flow rate: 80-100ml/min
- Warm-up Time: 30 minutes from cold start.
O2 Analysis
- Patent Number 4995256
- Type: Zirconia
- Range: 0 - 100%
- Response: <80 msec
- Accuracy: ± 0.03% 02
CO2 Analysis
- Type: NDIR
- Range: 0 - 10%
- Response: <130 msec
- Accuracy: ± 0.1% CO2
Size
- Height: 36 cm (14")
- Width: 36 cm (14")
- Depth: 36 cm (14")
- Weight: 12 kgs (26.5 lbs)
Power Requirements
|