Standard Features
Options
- Pulmonary Consult Interpretation Software
- Exercise Consult Interpretation Software
- Bronchial Provocation
- Networking & Information Management
- Database Query
- ExFVL (Exercise Flow Volume Loops)
Flow Device
- Bi-directional Pitot Tube Pneumotach
- Patent Number: #5,038,773 & #5,119,825
- Range: ± 18 L/sec
- Accuracy: ± 3% or 50 ml, whichever is greater
(meets or exceeds ATS/ERS clinical performance standards)
- Resolution: 8.64 ml/sec
- Deadspace: 39 ml
Gas Sample
- Patent Number 5,042,500
- Patented gas drying sample circuit
- Side stream sampling flow rate: 80-100ml./min
- Warm-up Time: 30 minutes from cold start.
Power Requirements

All
specifications are subject to change without notice.
Products may vary from those illustrated. Please
contact your Medical Graphics representative for
the latest information, pricing and product availability.
©
2005-2006 Medical Graphics Corporation. Ultima and
preVent are registered trademarks of Medical
Graphics Corporation.
|
Gas Requirements
- 100% Helium (10 - 28 psi)
- DLco mix (135 psi)
- 100% Oxygen (135 psi)
- Calibration Gas:
- 5% CO2, 12% O2, Bal N2
- 0% CO2, 21% O2, Bal N2
O2 Analysis
- Patent Number 4,995,256
- Type: Zirconia
- Range: 0 - 100%
- Response: (10-90%) <80 msec
- Accuracy: ± 0.03%
Arm Adjustments
- Vertical Extension: 43 cm (17”)
- Horizontal Extension: 35.5 cm (14”)
- Radius: 180º..
Size (Base)
- Height: 36 cm (14”)
- Width: 33 cm (13”)
- Depth: 36 cm (14”)
- Weight: 12 kgs (26.5 lbs)
Size (PF Module)
- Height: 15 cm (6”)
- Width: 33 cm (13”)
- Depth: 30 cm (11.5”)
- Weight: 9.3 kgs (20.5 lbs)
|